1 |
Power Supply Requirement: |
Single Phase 220VAC 10 % 50 Hz
Standard 15 Amp plug socket |
2 |
Tube Unite : |
Focal spot
0.6/1.8 mm |
3 |
Generator Rating Control : |
Microprocessor Based with Tactile Membrane
Switches |
4 |
Mobile Stand : |
|
|
C Orbital Movement |
115 (90 + 115) |
|
C Orbital Movement |
118 each side |
|
Motorized Vertical Movement |
440 mm |
|
Forward Movement |
200 mm |
|
Right to Left Wigwag Movement |
+/- 12.5 |
|
Index for C-Arm Rotating to get |
(90 + 90 + 90 + 90) AP/LAT Views |
5 |
Radiography Mode : |
|
|
KVP Output |
45 to 100 KV p (5KV step) |
|
MA Output |
20 Ma fixed |
|
MAS Range |
4 to 120 MAS |
6 |
Fluoroscopy Mode : |
|
|
KVp Output |
40 to 100 KVp |
|
MAS Output |
0.1 - 3.5 MA (Motorized steeples) |
|
Fluoro Timer |
0 to 5 min cumulative timer with buzzer |
|
Safety Precaution |
Temperature Sensor to cut off exposure
After overheating of x-ray tube head |
7 |
Image Intensifier : |
9" Triple field, high gain low I
Dose image intensifier |
8 |
T.V. Camera : |
High Sensitivity, low noise Sony CCD Camera |
9 |
Monitor : |
Two 19" Lcd monitors with high Density & low noise |
10 |
Memory : |
100 frames memory with full function remote control.
Real time zoom with vertical and Horizontal panning. |